BL9A High-intensity and Low-energy XAFS
2020/3/9
Techniques
XAFS(X-ray Absorption Fine Structure)
- Energy range: 2.1 – 15 keV (2.1 – 4 keV with SX mode)
- Transmission mode: QXAFS
- Fluorescent mode: Lytle detector, Multi Ge-SSD
- Conversion electron yield (CEY) mode: Available with SX mode
Other experimental with monochromatized X-ray
- Please contact us before experiment
Light source
PF Bending magnet #9
- Ring energy: 2.5 GeV
- Max. current: 450 mA (Usually Top-up operation)
- Critical energy: 4 keV
- Horizontal acceptance: 3 mrad
Beamline optics
Monochromator
- Si(111) Water-cooled double crystal
- Energy range: 2.1 – 15 keV
- Cam-drive (Kohzu system)
Focusing system
- Double Bent conical mirror (Rh coated)
for collimating and forcusing - Standard beam size: 0.5H x 0.3V mm
Others
- Double flat morror (Ni coated) for higher harmonics reduction
Focused and Monochromatized mode
- Energy range: 4 – 15 keV
- Energy resolution: ΔE/E~2×10-4 (9 keV)
- Photon flux at sample position: 6×1011 phs/s (7 keV)
Soft X-ray(SX) mode (Focused and Monochromatized)
- Energy range: 2.1 – 4 keV
- Photon flux at sample position: 6×1010 phs/s (2.5 keV)
Experimantal station
Experimental hutch
- Hatch size: 2.6L x 2.8W x 2.5H m
- Door size: 1.88W x 2.39H m
Standard setup
- Flat table size: 1500L x 1000W mm
- XAFS setup: Ionization chumber, Ge-SSD, Cryocooler
- SX setup: He filled CEY sample chamber
Beamline manager
- 阿部 仁 / ABE Hitoshi (KEK-PF)
hitoshi.abe[@]kek.jp